The DS‑36D series includes three benchmark desktop spectrophotometers: DS‑36D, DS‑37D, and DS‑39D. They use a differential spectral engine, dual optical paths, and 1nm grating technology, and feature a dual light source with pulsed xenon lamp + LED. The DS‑36D has a repeatability of ΔEab≤0.01, while the DS‑37D/39D achieve ΔEab≤0.005, with the DS‑39D’s inter-unit difference as low as 0.08. They also come with automatic fluorescence calibration, temperature and humidity compensation, support simultaneous SCI+SCE measurement, and are suitable for high-precision color management in coatings, plastics, textiles, and research labs.
$0.00
DetailSupports

Product Details

Model

DS-36D

DS-37D

DS-39D

Lighting/measuring

conditions


Reflection:d/8(diffuse illumination,8°direction reception)

SCI(Contains specular reflected light)/ SCE(not contain specular reflected light)measure at same time.

Compliance standards:CIE No.15、GB/T 3978、GB 2893、GB/T 18833、ISO7724/1、DIN5033 Teil7、

JIS Z8722 Condition C、ASTM E1164、ASTM-D1003-07

Transmission:d/0(diffuse illumination, vertical reception)

Sensor

Differential spectrum engine

Spectroscopic method

Concave grating

Integrating sphere diameter

152mm

Wavelength range

360nm-780nm

Wavelengthinterval

10nm

Reflectance measurement

range

0-200%,resolution 0.01%

Lighting source

Pulsed xenon lamps andLED

Ultraviolet measurement

IncludesUV、400nmcutoff、420nmcutoff、460nmcutoff

Measuring time

Single mode<2s

Lighting/measuring

calibers

Reflection:XLAV Φ25.4mm/Φ30mm;LAVΦ15mm/Φ18mm;MAVΦ8mm/Φ11mm;SAVΦ3mm/Φ6mm

Users can customize the calibre, and the calibre switch is automatically recognized

Transmission:Φ17mm/Φ25mm

Transmission measurement

specification

Sample height and thickness:height is not limited, thickness≤50mm

Repeatability *

ΔE*ab≤0.01,

Spectral reflection/transmittance ≤0.1%

ΔE*ab≤0.005,

Spectral reflection/transmittance ≤0.1%

Inter-Instrument

Agreement**

XLAV ΔE*ab 0.15

XLAV ΔE*ab 0.12

XLAV ΔE*ab 0.08

Long-term repeatability***

XLAV chroma value:standard deviation ΔE*ab within 0.01(under constant temperature conditions,the white correction plate is measured every hour within 24 hours)

Standard observer

2°Standard observer and 10°Standard observer

Viewing light source

A,B,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,U30,U35,DLF,NBF,TL83,TL84,ID50,ID65,

LED-B1,LED-B2,LED-B3,LED-B4,LED-B5,LED-BH1,LED-RGB1,LED-V1,LED-V2,LED-8

Language

Simplified Chinese,English, Traditional Chinese, Russian, Spanish, Portuguese, Japanese, Thai, Korean,German, French, Polish

Display content

Spectral data,Spectrogram, chromaticity data, chromaticity Data, chromaticity map, Pass/Fail judgment, Simulation color, Color evaluation, fog, liquid chromaticity, Color bias

Color space

CIE LAB,CIE LUV,LCh,Hunter Lab,Yxy,XYZ,Musell,s-RGB,βxy

Chroma index

WI(ASTM E313-20,ASTM E313-73,CIE,AATCC,Hunter,Taube,Berger Stensby),YI(ASTM D1925,ASTM E313-20,ASTM E313-73),Tint(ASTM E313-20),Isochromatic index Milm,color fastness,color changing fastness,ISO brightness,R457,A density,T density,

E density,M density,APHA/Hazen/Pt-Co((platinum cobalt index),Gardner(Gardner Index),Saybolt(SeibertIndex),Astm color,fog,

total transmittance, covering power, force, intensity

Color difference formula

ΔE*ab,ΔE*CH,ΔE*uv,ΔE*cmc,ΔE*94,ΔE*00,ΔEab(Hunter),555color tone classification

Storage

8GB

Screen size

7-inch capacitive touch screen

Operating system

Android

Power source

Dc regulated power supply

Operating temperature

and humidity

5~40°C,relative humidity 80%(35°C)below no condensation

Storage temperature

and humidity

-20~45°C,relative humidity 80%(35°C)below no condensation

Accessories

Power adapter,USB cable, transmission fixture, software U disk, black cavity, white board, green board,30mm aperture,

18mm aperture,11mm aperture,6mm aperture,support table,cuvette,damping handle

Optional accessories

Heating transmission jig(including control circuit),vertical bracket,pneumatic jacking rod(includingcontrol circuit),small sample holding accessories,reflection cupping plate,fiber test box,

micro transmission jig, rod box, European standard plug, American standard plug

Port

RS-232、USB、USB-B、Bluetooth

Camera positioning

Ultra HD camera(1400dpi)

Automatic calibration

√(Can greatly improve the long-term repeatability of the instrument)

Fluorescence calibration

√(Can automatically adjustthe UV intensity,and ensure that the value of the instrument is highly consistent with that of other imported instruments when measuring materials containing fluorescence)

Brightness calibration

√(Through the brightness calibration algorithm, the real color of ultra-dark samples is restored)

Others

The instrument can be measured sideways,up and down (using accessories); Automatic temperatureand humidity compensation function; PC side software save sample image function

※After instrument calibration, the white correction plate was measured 30 times at 5-second intervals to measure the standard deviation of the result in XLAV caliber

※※Based on 23°C,the average value of XLAV aperture measurement of 12 swatches of BCRA Series is measured

※※※XLAV chromatic value:standard deviation ΔE*ab within 0.1(0℃-40℃ Arbitrary temperature variation)







Options

The item code

                                                         Name

1.51.01.0069-0

12   colored bricks

1.51.01.0068-0

24   colored bricks

3.07.04.2019-0

Sintered-quartz   square cuvette (outer dimensions length 43 * width 43 * height 45 mm, wall   thickness 1.5 mm, optical path length 40 mm)

3.07.04.2006-0

810   quartz square cuvette (outer dimensions length 42.5 * width 12.5 * height   45mm, arm thickness 1.25mm, optical path length 10mm, clear surface 40mm)

3.07.04.2017-0

800   quartz cylindrical cuvette (size φ64*15mm, wall   thickness 2mm)

3.07.04.2005-0

800   quartz cylindrical cuvette (size 63*40mm, wall thickness 2mm)

3.07.04.2014-0

33mm   pathlength cuvette (outer dimensions 42.5 x 35.5 x 45 mm, wall thickness 1.25   mm, clear surface 40 mm, optical path 33 mm)






Instrument Demonstration


1.Demonstration of measuring paint test cards with the desktop spectrophotometer DS-36D



2.Demonstration of measuring paint test cards with the desktop spectrophotometer DS-37D



3.Demonstration of measuring paint test cards with the desktop spectrophotometer DS-39D



DS-39D Family Sheet English